AMETEK SLM Series RevC User Manual
Page 71

Sorensen SLM-Series DC Load
Remote Programming
M540071-01 Rev C
5-15
become null and void.
2. The minimum effective digit of the value is the fifth place after the decimal point.
3. If the value designated exceeds the specification of the electronic load, the mainframe will send
out the full-scale current value of the electronic load specification.
4. The frequency Range 1s 40.0~70.0Hz.
5. The engineering unit for frequency is Hz.
Example:
FREQ 50.0, to set frequency is 50.0Hz.
FREQ 60.0, to set frequency is 60.0Hz.
FREQ 0.1, to set frequency is 0.1Hz, that is to set DC.
VOLTAGE Limit
Purpose:
To set the upper/lower limit value of threshold voltage.
Command Syntax:
All Modules:
LIM:VOLT:{HIGH|LOW}{SP}{NR2}{;|NL}
Description:
This command is to set the upper/lower limit value of threshold voltage. When input voltage is lower
than the lower limit value or higher than the upper limit value, NG indicating light will come on to
indicate “NO GOOD."
Example:
LIM:VOLT:LOW 1.0 ; To set the lower limit value of threshold voltage to 1.0 V.
LIM:VOLT:HIGH 200.0 ; To set the upper Limit vale of threshold voltage to 200.0V.
CURRENT Limit
Purpose:
To set the upper/lower limit value of threshold current.
Command Syntax:
All Modules:
LIM:CURR{HIGH/LOW}{SP}{NR2}{;|NL}
Description:
This command is to set the lower limit value of threshold current. When load sink current is lower
than this lower limit value or higher than the upper limit value, NG indicating light will come on to
indicate “NO GOOD."
Example:
LIM:CURR:LOW:0.05 ; To set the lower limit value of threshold current to 0.05A.
LIM:CURR:HIGH:10.0 ; To the upper limit value of threshold current to 10.0A.
POWER Limit
Purpose:
To set the upper/lower limit value of threshold power (W).
Command Syntax:
SLM DC Modules:
LIM:POW:{HIGH|LOW}{SP}{NR2}{;|NL}