AMETEK SLM Series RevC User Manual
Page 80

Remote Programming
Sorensen SLM-Series DC Load
5-24
M540071-01 Rev C
Reset by CLER command.
Example:
SLM-60-30-150
FALL 0.050 MA
FALL 100.0 MA
setting fall time to 100.0 MA will change range then bit 1 will set to high.
bit 2 :invalid command
This bit set to high by accepted illegal command.
bit 3 :invalid operating
This bit set to high by accepted invalid operating command.
Example:
SLM-60-30-150
MODE CR
DYN ON
DYNAMIC function only supports CC MODE.
VOLTAGE METER
Purpose:
The reading of 4 1/2 digit voltage meter read back query command.
Command Syntax:
All Modules:
MEAS:VOLT?{NL}
Description:
MEAS:VOLT? returns the present 4 1/2 digital voltage meter reading. The returned data format is
shown in Table 4-2, the engineering unit is “V”.
CURRENT METER
Purpose:
The reading of 4 1/2 digit current meter read back query command.
Command Syntax:
All Modules:
MEAS:CURR?{NL}
Description:
MEAS:CURR? returns the present 4 1/2 digital current meter reading. The engineering unit is “A."
POWER METER
Purpose:
To read the value of Watt meter.
Command Syntax:
SLM Modules:
MEAS:POW?{;|NL}
Description:
MEAS:POW? Reads back the value of 4 digit of the Watt meter; unit is (W).
VA METER
Purpose:
To read the value of VA meter.