Operation

Sorensen SLM-Series DC Load

3-4

M540071-01 Rev C

9

LEVEL Key and Hi/Lo LED

LEVEL key toggles between the High or Low levels for programming the high and low

level load values for CC, CR, CV, CP when in those modes. (See Key 16 for setting
the Level values. (These High/Low level load values can be programmed while in
Static mode but not in Dynamic mode. See Key 10, DYN/STA, next).

LED indicates whether the Load value being programmed is High level or Low level:

◦ LED lit = high (HI) level
◦ LED not lit = low (Lo) level

Section 3.2.3 provides guidelines for programming the values.

10 DYN / STA Key and LED

DYN/STA key toggles between Dynamic and Static modes only when in Constant

Current mode or Constant Power mode. (When in Constant Resistance and Constant
Voltage modes, the SLM-series DC load module automatically adjusts to Static mode
because Dynamic mode is not available for either).

LED indicates whether the SLM-series DC load module is in Dynamic or Static mode:

◦ LED lit = Dynamic
◦ LED not lit = Static

11 SHORT Key and ON/OFF LED

The SHORT key toggles the short circuit of DC input terminal On and Off.

LED indicates the ON/OFF status of the short circuit

◦ LED lit = ON
◦ LED not lit = OFF

12 LIMIT Key and LED

The Limit key sequentially selects the following parameters to program their settings:

◦ OFF
◦ GO/NG Upper/Lower Voltage limit
◦ GO/NG Upper/Lower Current limit
◦ GO/NG Upper/Lower Power limit
◦ Vsense Auto/ON
◦ Load ON voltage
◦ Load OFF voltage
◦ Range: Auto/RangeII
◦ iStart
◦ iStop
◦ iStep
◦ Pstart
◦ Pstop
◦ Pstep
◦ V threshold