Remote Programming
Sorensen SLM-Series DC Load
5-12
M540071-01 Rev C
MODE CP is available in DC, single input loads only.
CLEAR status register
Purpose:
CLEar the PROT and ERR status byte registers.
Command Syntax:
All Modules:
CLER{NL}
Description:
CLER ; clear the PROT and ERR status byte register, the PROT and ERR status byte register will
indicate “0" after executing the CLER command.
STORE
Purpose:
STORE the load level and load status into the memory of the SL-series electronic LOAD.
Command Syntax:
SLM DC, SLD Modules: STOR{SP}{m[,n]}{;|NL}
SLM AC Modules
STOR{SP}{m}{;|ML}
Description:
Parameter m is 1~5 for 5 different states withSL-series electronic load module's load status and
load current into the non-volatile memory.
Parameter n is 1-30 for 30 memory bank for 150 (m*n) different state with DC electronic load
module's load status and load current into the EEPROM memory in the electronic loads.
Example:
STORE 1; store the AC electronic load module's load status and load current into the memory 1.
STORE 2,30; store the DC electronic load module's load status and load current into the memory
147.
RECALL
Purpose:
Recall the state of load level and status, is stored by the GPIB/RS232 STORe command.
Command Syntax:
SLM DC, SLD Modules: REC{SP}{m[,n]}{;|NL}
SLM AC Modules:
REC{SP}{m}{;|NL}
Description:
This command is used to recall the memory state, is stored into memory by the GPIB/RS232
STORe command, up to 5 states can be recalled for AC electronic load modules, and up to 150
states can be recalled for DC electronic load modules.
Example:
REC 1; Recall the state of load level and status that is stored in memory 1 by GPIB/RS232 STOR
command.
REC 147; Recall the state of load level and status that is stored in memory 147 by GPIB/RS232
STOR command.
SYNCHRONOUS ON/OFF
Purpose: