AMETEK SLM Series RevC User Manual
Page 64

Remote Programming
Sorensen SLM-Series DC Load
5-8
M540071-01 Rev C
Description:
CC:{LOW|HIGH}{SP}{NR2}{;|NL}
Sets the current level of SLM-series AC or DC Electronic Load modules.
CC:{SP}{NR2}{;|NL}
This command is used to set the load current level for CC static mode of SLD-series electronic load
module.
Note:
a. The load current data must include the decimal point; otherwise, this command will not
execute. The load current level can be programmed up to the sixth place after the
decimal point.
b. The HIGH level load current MUST be higher than the LOW level load current (and vice
versa) for proper dynamic waveform definition; if not, the SLM-series Electronic Load will
adjust and limit the programmed values to be equal. The adjustment matches the
second input value to the first input value.
This means that if the value for the LOW level is input first, and then the HIGH level
value is input as less than the programmed LOW level, the SLM-series load module will
adjust the HIGH level to be equal to the LOW level. If the value for the HIGH level is
input first and the LOW level value is input as higher than the programmed HIGH level,
the SLM-series load module will adjust the LOW level to be equal to the HIGH level.
c. If the programmed load current level is over the maximum rated specification, the full
scale current will be sent to the load module.
d. Engineering unit for load current is Amps.
e. Please refer to Appendices C, D and E for proper programming procedure of SL-series
electronic load modules.
Example:
CC:LOW 1.8 set LOW level load current to 1.8 A.
CC:HIGH 25.123456 set HIGH level load current to 25.123456 A.
RESISTANCE
Level
Purpose:
The load resistance setting in Constant Resistance mode.
Command Syntax:
All SLM Modules:
CR:{HIGH|LOW}{SP}{NR2}{;|NL}
SLD Modules
CR:{SP}{NR2}{;|NL}
Description:
CR:{HIGH|LOW}{SP}{NR2}{;|NL}
This command is used to set the LOW/HIGH load resistance level of SLM series AC and DC
electronic load module.
CR:{SP}{NR2}{;|NL}
This command is used to set the load resistance level of SLD-Series load module.
Note:
a. The load resistance data must include the decimal point; otherwise, this command will
not execute. The load resistance level can be programmed up to the sixth place after the