Sorensen SLM-Series DC Load

Operation

M540071-01 Rev C

3-3

6

Lower 5-Character Digital Meter (DM) and Associated LEDs

Multi-purpose display, depending on selected mode:

In PRESet OFF mode, displays:

◦ actual load current when SHORT status is OFF
◦ short circuit current when SHORT status is ON

In PRESet ON mode, displays values being programmed for:

◦ CC
◦ T

HIGH

/T

LOW

when in Dynamic mode

◦ RISE/FALL slew rates when in Dynamic mode

In LIMIT ON mode, displays:

◦ oVP (over voltage), oCP (over current), oPP (over power), or

oTP (over temperature) when in protection mode

◦ GO/NG low limit values for voltage, current, and power
◦ Auto or On setting for V

SENSE

control

◦ either Ld-on or Ld-off to indicate whether Load voltage setting mode is On or

Off

Associated LEDs indicate the displayed value's unit:

V = Value displayed is in Volts

A = Value displayed is in Amps

Ω = Value displayed is in Ohms

W = Value displayed is in Watts

ms = Value displayed is in milliseconds

A/

μs (or mA/μs) = Value displayed is in Amps per microsecond (or milliamps per

microsecond)

7

PRES Key and ON/WATT LED

PRES key sequentially selects Preset ON, Watt ON and Preset OFF

LEDs indicate which is selected at any given time:

◦ Preset On = LED is lit amber
◦ Watt On = LED is lit green
◦ Preset Off = LED is not lit

8

LOAD Key and ON/OFF LED

LOAD key toggles ON/OFF the DC load. (Does not affect the programmed settings).

LED indicates the status of the Load function:

◦ LED lit indicates the Load function is ON and ready to sink current from the DC

input.

◦ LED not lit indicates the Load function is OFF.