AMETEK SLM Series RevC User Manual

Page 79

background image

Sorensen SLM-Series DC Load

Remote Programming

M540071-01 Rev C

5-23

bit 7

bit 0

7

6

5

4

3

2

1

0

OPP

OTP

OVP

OCP

Figure 5-2 Protection Status Register

ERROR Status register

Purpose:
ERRor status register query command.

Command Syntax:
All Modules:

ERR?{;|NL}

Description:
ERR? returns the present error status; the status byte register summarizes all of the error status
events from all status registers. the following figure describes the status byte the happened on the
SLM-series electronic load. The ERR status byte register is cleared when a CLER command clears
all of the PROT and ERR status registers.

bit 7

bit 0

7

6

5

4

3

2

1

0

Limited

Range Changed

Invalid Command

Invalid Operating

Figure 5-3 Error Status Byte Register

bit 0:Limited
This bit set to high by Electronic Module Load setting command causing over range. Reset
by CLER command.

Example:

SLM-60-30-150
MODE CC
RANG 1

CURR:HIGH 20.0
setting current=20.0 A
actual setting current=3.071 A
bit 0 will set to high

bit 1:Range Changed
This bit set to high by Electronic Module Load setting command causing change range.