The connections for the rear panel RS-232 interface are shown below; Figure 5-1-A depicts the
connector wire diagram, and Figure 5-1-B depicts the connections using a standard RS-232 cable.
RS-232C
On SLM
Tx
Rx
RTS
CTS
.
.
.
Inside
SLM
Tx
Rx
RTS
CTS
DS
GR
DC
DTR
PC -232C
Tx
Rx
RTS
CTS
.
.
.
A
2
3
4
5
6
7
8
9
RS-232C
On SLM
Tx
Rx
RTS
CTS
.
.
.
Inside
SLM
Tx
Rx
RTS
CTS
DS
GR
DC
DTR
PC -232C
Tx
Rx
RTS
CTS
.
.
.
2
3
4
5
6
7
8
9
B
Figure 5-1 RS-232 Interface Diagram
The following RS-232 setting commands are channel-dependent commands except ”CHAN” which is a
channel-specific
command; therefore, for proper program execution ”CHAN” should be sent first, and
then the channel-dependent command.
For example:
Short ON of Channel 1 of SL-series Electronic Load module, the RS-232 programming
command is: CHAN 1;SHOR ON.
As with the GPIB commands, the following RS-232 commands with [GLOB:] option can set all the SL-
series Electronic load modules in the SLM mainframe to be active simultaneously. This feature can
greatly reduce the testing time and increase efficiency.
5.3
GPIB/RS-232C Command List
5.3.1
Command Syntax Abbreviations
SP
:Space, the ASCII code is 20 Hexadecimal.
;
:Semicolon, Program line terminator, the ASCII code is OA Hexadecimal.
NL
:New line, Program line terminator, the ASCII code is OA Hexadecimal.
N
:Integer from 1 to 8.
NR2
:Digits with decimal point. It can be accepted in the range and format of ##.#####.
Example:
30.12345, 5.0