AMETEK SLM Series RevC User Manual
Page 74

Remote Programming
Sorensen SLM-Series DC Load
5-18
M540071-01 Rev C
SLEW Rate
Purpose:
Set the load current slew rate of SLD-series electronic load module.
Command Syntax:
SLD Modules
SLEW{SP}{NR2}{;|NL}
Description:
Rise and Fall slew rate is the same for the SLEW command setting of SLD electronic load module.
The Slew rate has two ranges, and it follows the CC mode range change automatically: when CC
Dynamic mode is in Range 1, the slew rate is in Range 1, if CC Dynamic mode is in Range 2, then
Slew rate is in Range 2.
SHORT ON/OFF
Purpose:
Short the DC input of Electronic load module.
Command Syntax:
SLM DC, SLD Modules: [GLOB:]SHOR{SP}{0|1|OFF|ON}{NL}
Description:
This command applies the short across the input of the Electronic load. Executing SHOR does not
affect any programmed settings and the Electronic load will return to those settings when the short
is removed.
Example:
GLOB:SHOR ON ; set all load module load input to short state.
CHAN 2;SHOR OFF ; set channel l, 2 load module load input short open state.
SENSE ON/OFF
Purpose:
Set the voltage sense ON/OFF of Electronic load.
Command Syntax:
All Modules:
[GLOB:]SENS{SP}{0|1|OFF|ON}{NL}
Description:
Set the Vsense inputs Off or On. If Vsense is set to Off, voltage measurement is taken at the load
input terminals. If Vsense is set to On, voltage is measured at the Vsense input. On SLM-seris
DC loads, Vsense is auto/off or on. If it is set to auto/off, the module will measure the Vsense input
if the voltage is greater than 1V (on 60V models) or 3V (on 250 and 500V models).
Example:
GLOB:SENS ON set all load module V-sense input to sense ON state.
CHAN 2;SENS OFF set channel l, 2 load module V-sense input to sense OFF state.
DYNAMIC ON/OFF
Purpose:
Set DYNamic ON or OFF command.
Command Syntax:
SLM DC, SLD Modules: [GLOB:]DYN{SP}{0|1|OFF|ON}{NL}
Description: