Remote Programming

Sorensen SLM-Series DC Load

5-10

M540071-01 Rev C

d. Engineering unit for load current is Volts.
e. Please refer to Appendices C, D and E for proper programming procedure of SL-series

electronic load modules.

Example:
CV:LOW 3.0 set LOW level load voltage to 3.0 V.
CV:HIGH 45.123456 set HIGH level load voltage to 45.123456 V.

POWER Level

Purpose:
The load power setting in Constant Power mode.

Command Syntax:
SLM DC Modules:

CP:{HIGH|LOW}{SP}{NR2}{;|NL}

Description:
This command is used to set the load Power level of electronic load modules.

Note

: Mode CP is available in SLM-series DC loads only.

LOAD ON/OFF

Purpose:
Turn the Electronic load module input ON or OFF.

Command Syntax:

All Modules:

[GLOB:]LOAD{SP}{0FF|ON}{NL}

Description:
This command sets the Electronic load to sink current from power source. GLOB:LOAD ON All the
Electronic load modules in the SLM mainframe are ready to sink current from power source.

For Example :
GLOB:LOAD OFF ; All load modules in the SLM mainframe are at input OFF condition.
CHAN 3:LOAD ON ; Set the channel 3 load module to LOAD ON status, this load module is ready
to sink current from the power source.
CHAN 1:LOAD 0; Set the channel 1 load module to LOAD OFF.

LOAD

ON VOLTAGE Setting

Purpose:
The Load ON voltage setting (Initial is 1.0V) of DC electronic load modules.

Command Syntax:
All Modules:

LDON{SP}{NR2}{;|NL}

Description:
The Load On voltage can be adjusted by the LDON command. The Range 1s 0.1-25.0 V (Res. =
0.1V). The load will start to sink current if power source output voltage is higher than Load On
voltage.
For Example:

LDON 2.5; Set the Load On voltage to 2.5V, The load will start to sink current when the power
source output voltage is higher than 2.5V.

LOAD OFF VOLTAGE Setting

Purpose

: