2 typical connections – AMETEK SLM Series RevC User Manual
Page 26

Installation
Sorensen SLM-Series DC Load
2-4
M540071-01 Rev C
Plug Connectors and Hook Terminals
This method is recommended when input current is greater than 20A.
Plug Connectors and Direct Insertion
This method is also recommended when input current is greater than 20A.
2.2.2 Typical
Connections
A. Local Sense Connection
B. Remote Sense Connection
Figure 2-3 Typical Connections for SLM-Series DC Series Load Module
2.2.3
The Connection for a Multiple Output Power Supply
Rule for a multiple output power supply connection to the SLM-series DC electronic loads:
The potential of positive input (Red binding post) must be higher than the potential of
negative input (Black binding post) of SLM-series DC electronic load.
Figure 2-4 is an example of +5V, -5V, +12V, -12V and +24V six outputs power supply connected
to a SLM-series DC electronic load.
Figure 2-4 SLM-Series DC Plug-in Load and Multiple Output Power Supply Connection
- +