beautypg.com

2 query commands – AMETEK SLM Series RevC User Manual

Page 76

background image

Remote Programming

Sorensen SLM-Series DC Load

5-20

M540071-01 Rev C

5.4.2 Query

Commands

CHANNEL

Purpose:
The active channel query command.

Command Syntax:
System: CHAN?{NL}

Description:
CHAN? return the active channel, “1-4” indicate the active channel is channel “1-4”.

CURRENT Level

Purpose:
The Constant Current mode's load current level query command.

Command Syntax:
All SLM Modules:

CC:{HIGH|LOW}{?}{;|NL}

SLD Modules:

CC{?}{;|NL}

Description:
CURR:LOW? return the presently programmed low load current level in Constant Current mode of
SLM-series DC or AC electronic load module.
CURR? return the presently programmed load current level in Constant Current mode of SLD-
series electronic load module.

RESISTANCE Level

Purpose:
The Constant Resistance mode's load resistance level query command.

Command Syntax:
All SLM Modules:

CR:{HIGH|LOW}{?}{;|NL}

SLD Modules:

CR{?}{;|NL}

Description:
RES:LOW? returns the presently programmed low load resistance level in Constant Resistance
mode of SLM-series DC or AC electronic load module.
RES? returns the presently programmed load resistance level in Constant Resistance mode of
SLD-series electronic load module.
The engineering unit is “OHM”.

VOLTAGE Level

The Constant Voltage mode's load voltage level query command.

Command Syntax:
SLM DC Modules:

CV:{LOW|HIGH}?{NL}

SLD Modules:

CV:{NL}

Description:

CV:LOW? return the presently programmed low load voltage level in Constant Voltage mode of
SLM DC electronic load module.