MKIII OPERATOR'S MANUAL

82

OP_ST.nnn - Operating State File
The operating state file controls many of the operating parameters of the Brewer. The left column
lists the actual value written in the OP_ST.nnn file. The middle column is the BASIC variable name
used in the Brewer software to contain this value, and the right column is a description of the
value's meaning.

# Sample

SW
Variable

Explanation

1 046

NO$

Brewer ID number

2 \BDATA

DD$

Data Directory

3 ICFjjjyy

ICF$

instrument constants file

4 ZSFjjjyy

ZSF$

zenith sky coefficients file

5 DCFjjjyy

DCF$

dispersion constants file

6 UVRjjjyy

UVR$

UV response file

7 01

DA$

Current day

8 01

MO$

Current month

9 99

YE$

Current year

10 Saskatoon LO$

Location of Brewer instrument

11 52.108

L1$

Latitude of instrument

12 106.713

L2$

Longitude of instrument

13 960

L3$

Average climatic station pressure (millibars)

14 1.8

TE$

Voltage representation of Brewer temperature

15 215

NC%

Azimuth north correction

16 0

HC%

Zenith horizon correction

17 14689

SR%

Azimuth steps per revolution

18 1

Q1%

Zenith drive motor

19 1

Q2%

Azimuth drive motor

20 1

Q3%

Iris drive motor

21 1

Q4%

Filterwheel #1 drive motor

22 1

Q5%

Filterwheel #2 drive motor

23 1

Q6%

Clock board

24 1

Q7%

A/D board

25 1

Q8%

UVB port

26 1

Q9%

Filterwheel #3 drive motor

27 1

Q10%

New or old temperature circuit. Set to 1 for a new
temperature circuit and 0 for an old circuit.

28 1

Q11%

Second film polarizer

29 0

Q12%

Set to 1 to enable NOBREW operation

30 1

Q13%

Wide HG slit present. Always set this to 1.

31 1

Q14%

New Brewer electronics board

32 0

Q15%

Humidity Sensor

33 skc/menu DI$

Schedule or menu indicator

34 O

3

MDD$

Mode

35 UMKNO2 SK$

Schedule name