MKIII OPERATOR'S MANUAL
82
OP_ST.nnn - Operating State File
The operating state file controls many of the operating parameters of the Brewer. The left column
lists the actual value written in the OP_ST.nnn file. The middle column is the BASIC variable name
used in the Brewer software to contain this value, and the right column is a description of the
value's meaning.
# Sample
SW
Variable
Explanation
1 046
NO$
Brewer ID number
2 \BDATA
DD$
Data Directory
3 ICFjjjyy
ICF$
instrument constants file
4 ZSFjjjyy
ZSF$
zenith sky coefficients file
5 DCFjjjyy
DCF$
dispersion constants file
6 UVRjjjyy
UVR$
UV response file
7 01
DA$
Current day
8 01
MO$
Current month
9 99
YE$
Current year
10 Saskatoon LO$
Location of Brewer instrument
11 52.108
L1$
Latitude of instrument
12 106.713
L2$
Longitude of instrument
13 960
L3$
Average climatic station pressure (millibars)
14 1.8
TE$
Voltage representation of Brewer temperature
15 215
NC%
Azimuth north correction
16 0
HC%
Zenith horizon correction
17 14689
SR%
Azimuth steps per revolution
18 1
Q1%
Zenith drive motor
19 1
Q2%
Azimuth drive motor
20 1
Q3%
Iris drive motor
21 1
Q4%
Filterwheel #1 drive motor
22 1
Q5%
Filterwheel #2 drive motor
23 1
Q6%
Clock board
24 1
Q7%
A/D board
25 1
Q8%
UVB port
26 1
Q9%
Filterwheel #3 drive motor
27 1
Q10%
New or old temperature circuit. Set to 1 for a new
temperature circuit and 0 for an old circuit.
28 1
Q11%
Second film polarizer
29 0
Q12%
Set to 1 to enable NOBREW operation
30 1
Q13%
Wide HG slit present. Always set this to 1.
31 1
Q14%
New Brewer electronics board
32 0
Q15%
Humidity Sensor
33 skc/menu DI$
Schedule or menu indicator
34 O
3
MDD$
Mode
35 UMKNO2 SK$
Schedule name