Operational information – Kipp&Zonen Brewer MkIII User Manual
Page 126
MKIII OPERATOR’S MANUAL
123
19. OPERATIONAL INFORMATION
A variety of operational status values define the current state of the Brewer. Commands exist
to set and read each of these. They share a common syntax so the complete set are listed in
the table below:
Syntax
?
or
?
to read the status variable
!
or
!
to write the status variable
Response
? form only
Parameters
parameter / response
format
Meaning
identifier
parameter in question
identifier or
number
which of many
depends on
depends on
Possible Names
Format of
meaning of
ANALOG.NOW
[
10-bit integer
no
A/D output value in A/D units
BREAK.ABORT.TIME
float seconds
Yes, reset to
0.25 on tepid
resets
The elapsed time in seconds for a continuous
TTY line-break to abort command processing.
BREAK.RESET.TIME
float seconds
Yes, reset to 5.0
on tepid resets
The elapsed time in seconds for a continuous
TTY line-break to force a warm restart of the
Brewer.
BYTE.X [
]ANSI C integer format
(Hex on output)
yes
value of the relevant byte in external RAM
BYTE.F [
]ANSI C integer format
(Hex on output)
no
reads value from Flash memory
BYTE.C [
]ANSI C integer format
(Hex on output)
no
value of the relevant byte in CODE space
BYTE.D [
]ANSI C integer format
(Hex on output)
yes
value of the relevant byte in DATA space
DIGITAL.INPUT
[
OFF or ON
no
reads whether input is ON or OFF; logic
levels are transparent to the user.
DIGITAL.OUTPUT
[
OFF or ON
yes
digital output point ON or OFF with logic
levels transparent to the user.
ECHO.SUPPRESSION
ON/OFF
Suppresses TTY mode echoing when ON,
allows echoing when OFF.
HG.SWITCH
OFF or ON
yes
mercury lamp set OFF or ON
STD.SWITCH
OFF or ON
yes
standard lamp set OFF or ON
LAMP.POWER [
no
lamp power consumption
LAMP.STATE [
lamp state type
yes
state of lamp (OFF or ON)
MOISTURE
Grams/cu. m
no
Moisture content of air inside the Brewer
MOISTURE.HIGH
Grams/cu m
yes
Moisture content high (watermark)
MOTOR.ALLSTILL
TRUE or FALSE
no
True only if all the motors have stopped
moving
APPENDIX H COMPUTER / BREWER INTERFACE (TELETYPE)