19. OPERATIONAL INFORMATION

A variety of operational status values define the current state of the Brewer. Commands exist
to set and read each of these. They share a common syntax so the complete set are listed in
the table below:

Syntax

?

or

?[]

to read the status variable

!

or

![]

to write the status variable

Response

? form only

Parameters

parameter / response

format

Meaning

identifier

parameter in question

identifier or
number

which of many

depends on

depends on

Possible Names

Format of writeable?

meaning of

ANALOG.NOW
[]

10-bit integer

no

A/D output value in A/D units

BREAK.ABORT.TIME

float seconds

Yes, reset to
0.25 on tepid
resets

The elapsed time in seconds for a continuous
TTY line-break to abort command processing.

BREAK.RESET.TIME

float seconds

Yes, reset to 5.0
on tepid resets

The elapsed time in seconds for a continuous
TTY line-break to force a warm restart of the
Brewer.

BYTE.X [

]

ANSI C integer format
(Hex on output)

yes

value of the relevant byte in external RAM

BYTE.F [

]

ANSI C integer format
(Hex on output)

no

reads value from Flash memory

BYTE.C [

]

ANSI C integer format
(Hex on output)

no

value of the relevant byte in CODE space

BYTE.D [

]

ANSI C integer format
(Hex on output)

yes

value of the relevant byte in DATA space

DIGITAL.INPUT
[]

OFF or ON

no

reads whether input is ON or OFF; logic
levels are transparent to the user.

DIGITAL.OUTPUT
[]

OFF or ON

yes

digital output point ON or OFF with logic
levels transparent to the user.

ECHO.SUPPRESSION

ON/OFF

Suppresses TTY mode echoing when ON,
allows echoing when OFF.

HG.SWITCH

OFF or ON

yes

mercury lamp set OFF or ON

STD.SWITCH

OFF or ON

yes

standard lamp set OFF or ON

LAMP.POWER [] watts

no

lamp power consumption

LAMP.STATE []

lamp state type

yes

state of lamp (OFF or ON)

MOISTURE

Grams/cu. m

no

Moisture content of air inside the Brewer

MOISTURE.HIGH

Grams/cu m

yes

Moisture content high (watermark)

MOTOR.ALLSTILL

TRUE or FALSE

no

True only if all the motors have stopped
moving

APPENDIX H COMPUTER / BREWER INTERFACE (TELETYPE)