beautypg.com

Kipp&Zonen Brewer MkIII User Manual

Page 80

background image


MKIII OPERATOR’S MANUAL

77


FZOAVG.nnn

1

2

3

4

5

6

7

8

9

10

11

12 13 14 15

16 17

17897 326.3 +11 -3.1 +1.2 31/ 77 192 12 44.7 +13.5 -28.6 +12.8 7/ 7

508 8

17997 342.3 +3.3 -3.1 +1.0 13/ 56 179 13 0.0 +0.0 +0.0 +0.0 0/ 3

0

0

18097 323.6 +2.9 -2.9 +0.9 9/ 55 190 14 332.6 +0.8 -25.3 +6.8 2/ 5

556 12

1. Julian day
2. DS ozone

10. FZ ozone

3. Standard deviation

11. Standard deviation

4. DS SO

2

12. FZ SO

2

5. Standard deviation

13. Standard Deviation

6. Good Observations

14 Good Observations

7. Total Observations

15. Total Observations

8. Harmonic mean of mu

16. Harmonic mean of mu

9. Representative hour

17. Representative hour


RSOAVG.nnn - Run / Stop Log
Used to monitor the operation of the slitmask. The values of colums 2,4,5,6,7,8, and 9 should be
1±0.003

1

2

3

4

5

6

7

8

9

08391

1.0006

1.1667

1.0003

.9995

1.0005

.9999

.9991

.9997

08491

1.0020

.9024

.9989

.9994

1.0001

.9992

.9998

.9996


1.dddyy

2-9 run/stop ratio for slit mask positions 0 through 7.


SCOAVG.nnn - Sun Scan Average File

1

2

3

4

5

6

7

8

9

07590

22

1.971933

1

133.6

440.1

131.7

-.3

135

07991

21

1.889433

2

134.5

418.1

131.3

.1

137

08791

30

1.804133

2

136.9

404.7

133.6

-.6

135

1. dddyy

6. Maximum Ozone

2. temperature

7. Minimum step of scan

3. airmass

8, Minimum SO

2

4. neutral density filter used

9. HG calibration point

5. maximum step of scan

SLOAVG.nnn - Standard Lamp Log
Used to monitor the stability of the Brewer’s ozone measureing stability..

1

2 3 4

5

6

7

8

9

10

11

12 13 14 15 16 17 18

06791 14 32 5 1705 815 -29 -702 3952 2023 581925 1 1 1 2 4 2 10848
06891 24 35 4 1707 815 -29 -703 3957 2026 572599 4 2 2 1 1 1 6619
06991 25 31 2 1706 815 -30 -700 3948 2021 576789 1 1 0 1 2 1 6873

1. dddyy

7. Ratio 3

2. low temperature of the SL test

8. Ratio 4

3. high temperature of the SL test

9. Ratio 5

4. number of sl tests

10. Ratio 6

5. ratio 1

11. Lamp intensity

6. ratio 2

12-18. Standard Deviations

APPENDIX A BREWER DATA FILES