Kipp&Zonen Brewer MkIII User Manual
Page 78
MKIII OPERATOR’S MANUAL
75
H2OAVG.nnn Humidity Log File
The file is updated during the AP command at the End-of-Day.
1
2
3
4
5
6
29098
23.939
21.186
24.793
3.60
19.4
29198
25.648
25.458
25.078
2.68
11.3
1. Julian day (jjjyy)
2. Temperature at the PMT (°C)
3. ‘Fan’ Temperature in °C - used in the absolute humidity calculation.
4. Temp of base plate (°C)
5. Moisture measured in grams of water per cubic meter of air.
6. Relative Humidity (%)
HGOAVG.nnn
1
2
3
4
08391
2592
33
18
08491
2330
25
20
08591
2483
34
22
1. Julian Day (jjjyy)
2 maximum lamp intensity of the days scans
3. high Brewer temperature of the days scans
4. number of scans.
MIOAVG.nnn - Micrometer Log
A new entry is generated each time the FR routine is run.
1
2
3
4
5
6
7
33898 o
3
28
286
-2
0
0
0
33998 o
3
27
286
0
0
0
0
34098 o
3
26
286
0
0
2
0
1. Julian day followed by O
3
to indicate this is an ozone operation
2. Temperature at time of test
3. wavelength offset step number
4. Micrometer #1 steps from operating position to zero sensor.
5. Micrometer #1 offset constant from the ICF file.
6. Micrometer #2 steps from operating position to zero sensor.
7. Micrometer #2 offset constant from the ICF file.
APPENDIX A BREWER DATA FILES