beautypg.com

Kipp&Zonen Brewer MkIII User Manual

Page 13

background image



MKIII OPERATOR'S MANUAL

10





























Figure 2.2: View of Brewer with Covers Removed


2.1

SPECTROPHOTOMETER



Table 2.1: Spectrophotometer Specifications

UV wavelengths:

‘ozone’ wavelengths (nm): 303.2 (Hg slit) , 306.3, 310.1,
313.5, 316.8, 320.1

Mercury-calibration

302.15 nm

Resolution

0.6 nm in UV

Stability

±0.01 nm (over full temperature range)

Precision

0.006 ±0.002 nm

Measurement range

286.5 nm to 363.0 nm

Exit-slit mask cycling

0.12 sec/slit, 1.6 sec for full cycle

O

3

Measurement accuracy

±1% (for direct-sun total ozone)

Ambient Operating temperature
range (for more information see
appendix F, Thermal Test)

0°C to +40°C (no heater)
-20C to +40°C (with heater option)
-50°C to +40°C (with complete cold weather kit)

Physical dimensions (external
weatherproof container)

Size: 70 by 46 by 34 cm
Weight: 34 kg (75 lbs)

Power requirements
Brewer and Tracker

90 to 264 VAC
47 to 440 Hz