MKIII OPERATOR'S MANUAL
128
Bad command string. Command string buffer cleared.
An illegal command was (not) processed. The subsequent message indicates the command.
Bad digital output setting. Only ON/OFF allowed.
(Self-explanitory)
Checksum bad in low level command.
During initialization, loadmode or opmode a message with a bad checksum was received.
This is normal if an initialization starts while a packet is being sent and is of no consequence
at that time. At other times it indicates a noisy communication line.
Command not accepted while another in progress.
Please notify KIPP & ZONEN <[email protected]> of the circumstances in which
this message was generated.
Config signature did not write properly.
The configuration was not correctly written to Flash memory. Possibly the Flash memory
chip has failed.
Flash config checksum error.
The data in the configuration is corrupted. Possibly the Flash memory chip has failed.
Flash config memory full.
Configuration can be saved to Flash memory(see SAVE command) only four times. Reload
the firmware.
Lamp
Only ON and OFF are valid settings for a lamp.
MAIN.C : motor initialization timed out.
Motor initialization failed to complete in the allotted time (RESET.TIME.OUT). Operation
proceeds with the motors which did initialize.
Motor
An attempt was made to initialize a motor but the motion which should have removed the
blocker from a reference sensor did not do so.
Motor
Motor has been reset to its reference position.
Warning that a motor encountered an end of travel limit. This indicates that for some time
prior to this message the motor was not in position.
Motor
An attempt was made to move a motor which has been configured to a MOTOR.CLASS of
NOMOTOR.
Motor
Error communicating with the controller for the given motor. If this error persists it indicates a
failure in the instrument electronics.
Motor
Please notify KIPP & ZONEN <[email protected]> of the circumstances in which
this message was generated.