MKIII OPERATOR’S MANUAL

117

Parameters

The are photon counts in the range 0 to 16777215, one for each slitmask position
measured in the most recent R command.

Example

See section on the R command.

10. PMT

Reads a count using the PMT. This command requires configuration variables.

Syntax

PMT

Response

The count value (decimal number using ASCII characters) is right justified and padded
with spaces in a nine-character string.


Parameters

Parameter /
response

format

meaning

0 to 2

24

-1

the PMT count taken

11. ** R

Measure the light intensity. This command requires configuration variables. The
configuration variable sets up the slit positions to correspond to a motor step position.

Syntax

R,,,

Parameters

may take values from 0 to 7, takes on values from to 7, and
takes on values from 1 to 255. In response the Brewer measures the light intensity for
each of the wavelength positions thru by running the slitmask from to
and back accumulating the counts for each separate position. This counts are
accumulated for repetitions of this back-and-forth scan. Every R command
zeroes the count accumulators for all slitmask positions. See the notes on the
configuration variable SLIT.PACE in section 5.19 Configuration Parameters.

If there are no parameters specified, the parameters from the previous R command
are used. If no previous R command parameters had been specified (as would be the
case after a warm start, for example) all of the parameters are assumed to be zero
and no scan is performed. In essence, all that happens is the count accumulators are
zeroed. The correspondence between or and slitmask position is given in
the following table:

Slitmask position

meaning

0

Hg calibration: 302.1 nm

1

Dark Count

2

λ1 306.2 nm

3

λ2 310.0 nm

4

λ3 313.5 nm

5

λ4 316.8 nm

6

λ5 320.0 nm

7

λ2 & λ4 for Deadtime test