** r – Kipp&Zonen Brewer MkIII User Manual
Page 120
MKIII OPERATOR’S MANUAL
117
Parameters
The
measured in the most recent R command.
Example
See section on the R command.
10. PMT
Reads a count using the PMT. This command requires configuration variables.
Syntax
PMT
Response
The count value (decimal number using ASCII characters) is right justified and padded
with spaces in a nine-character string.
Parameters
Parameter /
response
format
meaning
0 to 2
24
-1
the PMT count taken
11. ** R
Measure the light intensity. This command requires configuration variables. The
configuration variable sets up the slit positions to correspond to a motor step position.
Syntax
R,
Parameters
takes on values from 1 to 255. In response the Brewer measures the light intensity for
each of the wavelength positions
accumulated for
zeroes the count accumulators for all slitmask positions. See the notes on the
configuration variable SLIT.PACE in section 5.19 Configuration Parameters.
If there are no parameters specified, the parameters from the previous R command
are used. If no previous R command parameters had been specified (as would be the
case after a warm start, for example) all of the parameters are assumed to be zero
and no scan is performed. In essence, all that happens is the count accumulators are
zeroed. The correspondence between
the following table:
Slitmask position
meaning
0
Hg calibration: 302.1 nm
1
Dark Count
2
λ1 306.2 nm
3
λ2 310.0 nm
4
λ3 313.5 nm
5
λ4 316.8 nm
6
λ5 320.0 nm
7
λ2 & λ4 for Deadtime test