Kipp&Zonen Brewer MkIII User Manual
Page 75
MKIII OPERATOR'S MANUAL
72
A/D Values: -- A/D values are not stored in files (except HV and +5v and SL current), but they are
very important in assessing the health of the Brewer, so they are included here for reference.
A/D Values for nov 15/98 at 17:03:05 for instrument number 159
Channel#
Name
Value
Value
(Lamps off)
(Lamps on)
0
PMT temp (deg C)
28.50
28.50
1
Fan temp (deg C)
27.83
27.83
2
Base temp (deg C)
28.21
28.21
3
H.T. voltage (V)
1487.05 1487.05
4
+12V power supply (V)
11.91
11.91
5
+ 5V power supply (V)
4.97
4.97
6
-12V power supply (V)
-11.99
-11.99
7
+24V power supply (V)
24.48
24.48
8
Rate meter (V)
0.00
0.00
9
Below Spectro temp (C)
27.07
27.07
10
Window area temp (C)
28.50
28.50
11
External temp (deg C)
27.55
27.55
12
+ 5V ss (V)
5.00
5.00
13
- 5V ss (V)
-4.91
-4.91
14
Std lamp current (A)
0.00
1.57
15
Std lamp voltage (V)
0.00
10.00
16
Mer lamp current (A)
0.00
0.36
17
Mer lamp voltage (V)
0.00
11.46
18
External 1 (V)
0.07
1.36
19
External 2 (V)
0.05
0.70
20
External 3 (V)
1.19
1.19
21
Humidity (g/m3)
3.74
3.74
22
External 4 (V)
0.02
0.12
23
External 5 (V)
0.03
0.13