MKIII OPERATOR'S MANUAL
66
HG Calibration Data
Example Name
Hg
Type of measurement
12:10:22
Time of measurement
.9995
Correlation value
287.1829 Calculated micrometer step
287
Micrometer set to this step #
190255
Peak intensity of the HG scan
28
Temperature (deg. C)
Correlation Value: - The corelation between the stored and measured spectra.
- Calculated Micrometer Step number
- The micrometer position of the HG peak
Standard Lamp Test Data
#
Example
Name
1
Sl
type of measurement
2
A
Filter
3
0
ND filter position of filterwheel #2 (in steps)
4
737.41
time – minutes since 00:00 hrs
5
0
Lower slit mask position (1=dark)
6
6
Upper slit mask position (6=slit 5)
7
20
# of cycles
8
625382
raw counts wavelength #0
9
43
dark count
10 644575
raw counts wavelength #1
11 710539
raw counts wavelength #2
12 839228
raw counts wavelength #3
13 914419
raw counts wavelength #4
14 981043
raw counts wavelength #5
15 Rat
ratios header
16 1543
single ratio #1 MS(4)
17 1116
single ratio #2 MS(5)
18 385
single ratio #3 MS(6)
19 300
single ratio #4 MS(7)
These lines are repeated for a total of 7 sets of data per SL test. This data is then averaged in the
following summary.