Kipp&Zonen Brewer MkIII User Manual
Page 76
MKIII OPERATOR’S MANUAL
73
PO Values: -- The PO command generates the following.
MKIII BREWER INSTRUMENT #159
----------------------------
01-05-1998 17:02:42
***************************************************************************
* Ozone Values * 1 * 2 * 3 * 4 * 5 * hg *
***************************************************************************
Wavelength(nm) * 306.289 310.035 313.486 316.787 319.978 303.184
Temp. Coeff * 0.0000 -0.2473 -0.6914 -0.6902 -1.2794 0.0000
Disp. Coeff #1 * 2856.960 2896.561 2933.527 2968.578 3003.310 823.907
Disp. Coeff #2 * 0.076746 0.076004 0.075101 0.074407 0.073260
0.077476
Disp. Coeff #3 *-0.725E-6 -0.739E-6 -0.734E-6 -0.751E-6 -0.707E-6
-0.726E-6
***************************************************************************
ETC Values : O
3
= 1690 ; SO
2
= 215
O
3
Absn Coeffs : O
3
= 0.3446 ; SO
2
= 1.1533
SO
2
Absn Coeffs : O
3
= 0 ; SO
2
= 2.3500
Micrometer steps/deg = 0.00 WL cal step number = 286
Micrometer Zero = 2469 Umkehr offset = 1688
Iris Open Steps = 250 Zenith steps/rev = 2972
Micrometer 1 Offset O
3
= 0 Micrometer 2 Offset = 0
Grating Slope = 1 Grating Intercept = 10
Filterwheel 3 Offset O
3
= 242
Dead Time(ns) = 40 Buffer Delay(s) = 0.6
Zenith UVB Dome Position = 2223
Note: Faster Processors May Require a Longer Buffer Delay
(Typically 0.2 to 0.8 Seconds)
APPENDIX A BREWER DATA FILES