beautypg.com

Kipp&Zonen Brewer MkIII User Manual

Page 128

background image


MKIII OPERATOR’S MANUAL

125

Examples

?MOTOR.POS[IRIS]

reports the current position in half-step units of the iris motor.

!SUPPLY.VOLTAGE.HIGH[+5V] 5.00

resets the upper voltage watermark to the nominal value for the +5 volt supply.

Motor positions for instrument control

Motor # & Name

Step #

Position

Command String

1: Zenith Prism

0

pointing at standard lamp (internal)

M,1,0

1408

pointing at zenith sky (external)

M,1,1408

2112

pointing at UVB port (external)

M,1,2112

2: AzimuthTracker

0

reference direction (North)

M,2,0

14670

reference direction, 1 full turn CW from step
#0

M,2,14670

3: Iris

0

iris fully closed

M,3,0

75 or 250 iris fully open

M,3,75 or M,3,250

4: Filterwheel #1

320

0: film polarizer (horizontal)

M,4,320

256

1: quartz diffuser (translucent)

M,4,256

192

2: blocked aperture (opaque)

M,4,192

128

3: clear aperture (transparent)

M,4,128

64

4: quartz diffuser; ND of f=2.0 (translucent)

M,4,64

0

5: clear aperture (transparent)

M,4,0

5: Filterwheel #2

0

0: f = 0

M,5,0

64

1: f = 0.5

M,5,64

128

2: f = 1.0 f = neutral density factor

M,5,128

192

3: f = 1.5[Attenuation = 10

f

]

M,5,192

256

4: f = 2.0

M,5,256

320

5: f = 2.5

M,5,320

10: Micrometer #1 Approx 290 calibrated micrometer setting

M,10,xxxx

9: Micrometer #2

(nominal setting is 6±1.5mm; there are

M,9,xxxx

576 steps/mm)

(wavelength change of 0.006nm/step)

(positive steps increase wavelength and

decrease micrometer setting)

11: Slit Mask

0

0: slit 0 (HG) 303.2 – 426.4 nm

M,11,xxxx

2

1: dark count -----------------------

NOTE: R,0,6,2;O;A

4

2: slit 1 306.3 – 431.4 nm

gives a real time

6

3: slit 2 310.1 - 437.3 nm

listing of the registers

8

4: slit 3 313.5 - 442.8 nm

from 0 to 6 press

10

5: slit 4 316.8 - 448.1 nm

<> to stop.

12

6: slit 5 320.1 - 453.2 nm

14

7: dead time -----------------------


APPENDIX H COMPUTER / BREWER INTERFACE (TELETYPE)