Kipp&Zonen Brewer MkIII User Manual
Page 128
MKIII OPERATOR’S MANUAL
125
Examples
?MOTOR.POS[IRIS]
reports the current position in half-step units of the iris motor.
!SUPPLY.VOLTAGE.HIGH[+5V] 5.00
resets the upper voltage watermark to the nominal value for the +5 volt supply.
Motor positions for instrument control
Motor # & Name
Step #
Position
Command String
1: Zenith Prism
0
pointing at standard lamp (internal)
M,1,0
1408
pointing at zenith sky (external)
M,1,1408
2112
pointing at UVB port (external)
M,1,2112
2: AzimuthTracker
0
reference direction (North)
M,2,0
14670
reference direction, 1 full turn CW from step
#0
M,2,14670
3: Iris
0
iris fully closed
M,3,0
75 or 250 iris fully open
M,3,75 or M,3,250
4: Filterwheel #1
320
0: film polarizer (horizontal)
M,4,320
256
1: quartz diffuser (translucent)
M,4,256
192
2: blocked aperture (opaque)
M,4,192
128
3: clear aperture (transparent)
M,4,128
64
4: quartz diffuser; ND of f=2.0 (translucent)
M,4,64
0
5: clear aperture (transparent)
M,4,0
5: Filterwheel #2
0
0: f = 0
M,5,0
64
1: f = 0.5
M,5,64
128
2: f = 1.0 f = neutral density factor
M,5,128
192
3: f = 1.5[Attenuation = 10
f
]
M,5,192
256
4: f = 2.0
M,5,256
320
5: f = 2.5
M,5,320
10: Micrometer #1 Approx 290 calibrated micrometer setting
M,10,xxxx
9: Micrometer #2
(nominal setting is 6±1.5mm; there are
M,9,xxxx
576 steps/mm)
(wavelength change of 0.006nm/step)
(positive steps increase wavelength and
decrease micrometer setting)
11: Slit Mask
0
0: slit 0 (HG) 303.2 – 426.4 nm
M,11,xxxx
2
1: dark count -----------------------
NOTE: R,0,6,2;O;A
4
2: slit 1 306.3 – 431.4 nm
gives a real time
6
3: slit 2 310.1 - 437.3 nm
listing of the registers
8
4: slit 3 313.5 - 442.8 nm
from 0 to 6 press
10
5: slit 4 316.8 - 448.1 nm
<
12
6: slit 5 320.1 - 453.2 nm
14
7: dead time -----------------------
APPENDIX H COMPUTER / BREWER INTERFACE (TELETYPE)