Average files: ???AVG.nnn and OZOAVGYY.nnn

The Brewer is a ‘statistical’ instrument, and instantaneous deviations from the norm are not
uncommon. The purpose of average files is to provide a daily value for a specific measurement or
test result. It is normal to plot the data in average files vs time (days) to observe trends in data and
test results.


APOAVG.nnn - Analog Printout Log
The values should be monitored for power supply stability.

1

2

3

4

08591

1566.80

5.17

1.61

08691

1555.86

5.17

1.61

1. Julian Day (dddyy)
2. H.T. voltage
3. +5V
4. standard lamp current



DTOAVG.nnn - Dead Time Log

1

2

3

04991

41.66

43.375

05991

41.69

43.276

1. Julian Day (dddyy)
2. dead time for high intensities
3. dead time for low intensities



FMOAVG.nnn - FM Average

1

2

3

4

5

6

7

8

9

05791

425.9

+4.8

3152.5

-0.2

1.3

189001.4

4

33

05891

403.1

+5.3

2909.1

2.6

4.8

1788953

3

34

1. Julian Day (dddyy)
2. daily mean ozone
3. standard deviation of daily mean ozone
4. ETC
5. daily mean SO

2

6. standard deviation of daily mean SO

2

.

7. ETC,
8. number of good observations
9. number of total observations.