MKIII OPERATOR'S MANUAL

120

17. V

Set the baud rate and the flag which controls echoing.

Syntax

V,[,]

Parameters

Parameter

format

Meaning

byte

one of 30, 60, 120, 240, 480 or 960. The approximate
number of characters per second. Baud rate set to
10*

Boolean

on to suppress character echoing

Examples

V,120
The prompt ("->") after this command will be sent at 1200 baud

V,960,1

The prompt ("->") after this command will be sent at 9600 baud and character echoing will
be suppressed.

18. CONFIGURATION PARAMETERS

A variety of configuration parameters define the operation of the Brewer. Commands
exist to set and read each of these. They share a common syntax so the complete set
are listed in the table below:

Syntax

? or
?[]to read the configuration variable
! or
![] to write the configuration variable

Response

The >

Parameters

Parameter / response

Format

meaning

Identifier

starting address within the space

Identifier or number

which element of a vector of such
values

Depends on

depends on