MKIII OPERATOR’S MANUAL

129

Motor : Motor lost.

The motor controller and the instrument controller are out of step. This error will be corrected
next time the moter is moved. If this error persists for the given motor, please notify KIPP &
ZONEN <[email protected]> of the circumstances in which this message was
generated.

Motor : Move attempted outside of logical bounds

An attempt was made to move the given motor out of bounds. If this error persists, please
notify KIPP & ZONEN <[email protected]> of the circumstances in which this
message was generated.

Motor : Movement attempted with too large a jerk

Please notify KIPP & ZONEN <[email protected]> of the circumstances in which
this message was generated.

Motor : Movement attempted with too large an acceleration

Please notify KIPP & ZONEN <[email protected]> of the circumstances in which
this message was generated.

Motor : Movement attempted with too large a velocity

If this error persists for the given motor, please notify KIPP & ZONEN
<[email protected]> of the circumstances in which this message was generated.

Motor : time allotted ( seconds) for motor movement has expired. Motor
stopped.

The value comes from the MOTOR.TIME.OUT configuration parameter.

New value set for watermark : .

The indicated watermark value has been changed to . The possible watermarks
are identified in Appendix H.

No Flash config segment in Flash to copy into RAM upon initialization

Attempt to initialize when no configuration information had been loaded yet. This message is
normal as part of the process of loading new firmware.

Parallel command operation not supported:

A command was issued which requested that two operations run in parallel when that is not
legal. The subsequent two messages indicate that the command is bad and present the
command. The following commands cannot start while another command is still in operation:
I, PMT, R, SAVE, STEPS, USECONFIG

PMT counter failure

The two independent counters on the PMT did not match count rates (differed by more than
2).Counter 2 has been disabled until explicitly re-enabled or until the instrument reinitializes.

APPENDIX I FIRMWARE LOG