beautypg.com

Kipp&Zonen Brewer MkIII User Manual

Page 73

background image



MKIII OPERATOR'S MANUAL

70

16 625382

raw counts wavelength #0

17 120

dark count

18 97706

raw counts wavelength #1

19 294523

raw counts wavelength #2

20 759956

raw counts wavelength #3

21 1087648

raw counts wavelength #4

22 962462

raw counts wavelength #5

23 rat

Ratio

24 105525

Single ratio #1 MS(4)

25 5722

Single ratio #2 MS(5)

26 1572

Single ratio #3 MS(6)

27 -574

Single ratio #4 MS(7)


S Files: SJJJYY.nnn - ‘S’, or Summary files contain summary information from each test and
measurement contained within the B file for the same day
The data contained in the S file is essentially the same as the “summary” line of the B file.
SJJJYY.nnn files start with a “tc” and the 5 temperature coefficients which were in the instruments
constants file when the SUM command was issued. The SUM command is normally performed as
the first command in the ED end of day sequence. Each item in a S-file is separated with a carriage
return and line feed, the end of the summary file is marked with “ef”.

UV Files UVJJJYY.nnn
- ‘UV’ files contain data data from UV scans, Including UV, UX, and UA

# Example

Name

1 UX integration time is

0.2294 seconds per cycle

(header for each scan).

2 dt 4.3E-08

Dead time

3 cy 1

# of cycles

4 Dh

Data header

5 26

Day

6 04

Month

7 92

Year

8 Saskatoon

Location

9 52.108

Latitude

10 106.713

Longitude

11 2.56

Temperature (volts)

12 Pr

Pressure header

12 960 dark

Pressure (mb) and dark count header

14 1

Dark count

15 825.9

Time (GMT) in minutes for 1st λ

16 2865

Wavelength for 1st λ

17 1263

Micrometer step # for 1st λ

18 11

raw counts for 1st λ

19

Lines 15 to 18 are repeated for each λ


In UX scans, lines 15 to 18 are repeated for wavelengths 286.5nm to 363.0nm in 0.5nm
increments. In UV measurements the scan is from 290.0nm to 325.0nm then back to 290.0nm in
0.5nm increments and the number of cycles is 4 and counts for 4 cycles are recorded in the UV file.

Umkehr Files: UJJJYY.nnn -- As part of the End-of-Day process the Umkehr data in the B file is
transferred into a U file which contains both morning and evening data if it has been collected.