Kipp&Zonen Brewer MkIII User Manual
Page 73
MKIII OPERATOR'S MANUAL
70
16 625382
raw counts wavelength #0
17 120
dark count
18 97706
raw counts wavelength #1
19 294523
raw counts wavelength #2
20 759956
raw counts wavelength #3
21 1087648
raw counts wavelength #4
22 962462
raw counts wavelength #5
23 rat
Ratio
24 105525
Single ratio #1 MS(4)
25 5722
Single ratio #2 MS(5)
26 1572
Single ratio #3 MS(6)
27 -574
Single ratio #4 MS(7)
S Files: SJJJYY.nnn - ‘S’, or Summary files contain summary information from each test and
measurement contained within the B file for the same day
The data contained in the S file is essentially the same as the “summary” line of the B file.
SJJJYY.nnn files start with a “tc” and the 5 temperature coefficients which were in the instruments
constants file when the SUM command was issued. The SUM command is normally performed as
the first command in the ED end of day sequence. Each item in a S-file is separated with a carriage
return and line feed, the end of the summary file is marked with “ef”.
UV Files UVJJJYY.nnn - ‘UV’ files contain data data from UV scans, Including UV, UX, and UA
# Example
Name
1 UX integration time is
0.2294 seconds per cycle
(header for each scan).
2 dt 4.3E-08
Dead time
3 cy 1
# of cycles
4 Dh
Data header
5 26
Day
6 04
Month
7 92
Year
8 Saskatoon
Location
9 52.108
Latitude
10 106.713
Longitude
11 2.56
Temperature (volts)
12 Pr
Pressure header
12 960 dark
Pressure (mb) and dark count header
14 1
Dark count
15 825.9
Time (GMT) in minutes for 1st λ
16 2865
Wavelength for 1st λ
17 1263
Micrometer step # for 1st λ
18 11
raw counts for 1st λ
19
Lines 15 to 18 are repeated for each λ
In UX scans, lines 15 to 18 are repeated for wavelengths 286.5nm to 363.0nm in 0.5nm
increments. In UV measurements the scan is from 290.0nm to 325.0nm then back to 290.0nm in
0.5nm increments and the number of cycles is 4 and counts for 4 cycles are recorded in the UV file.
Umkehr Files: UJJJYY.nnn -- As part of the End-of-Day process the Umkehr data in the B file is
transferred into a U file which contains both morning and evening data if it has been collected.