SCPI Commands Reference

146

PSO-200

Product-Specific Commands—Description

Notes

When averaging is on and the number of scans is

less than or equal to the Avg Number setting, a

linear average is calculated. After the scan count

exceeds the Avg Number setting, if more burst

has to be acquired.

If Average Mode is Exponential then new results

are averaged in exponentially. In other words,

each succeeding average will be the weighted

sum of the previous average, weighted by (N–

1)/N, and the new measurement, weighted by

1/N, where N is the Average Number setting.

If Average Mode is Last, then average values will

be computed when the acquisition will stop.

The last N bursts will be used to compute the

averages.

State is saved in instrument.

See Also

SENSe:DDEMod:AVERage:COUNt 10

:SENSe:DDEMod:AVERage:TCONtrol