SCPI Command Reference

Optical Spectrum Analyzer

505

Product-Specific Commands—Description

:SENSe[1..n]:WAVelength[:RANGe]:

STOP

Description

This command sets the stop wavelength for

acquisition range.

For this command to be accepted, the OSA

module must be in Ready state with no

acquisition in progress.

At *RST, this value is set to 1565.0 nm (C-band

upper limit).

Syntax

:SENSe[1..n]:WAVelength[:RANGe]:STOP

NM]>|MAXimum|MINimum

Parameter(s)

Stop:
The program data syntax for is defined

as a element followed by an

optional element.

The allowed

element is NM. The special forms

MINimum and MAXimum are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

Maximum wavelength for acquisition range.

Range is device-dependent.