beautypg.com

EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual

Page 500

background image

SCPI Command Reference

492

FTB-5240S/5240BP

Product-Specific Commands—Description

:MMEMory[1..n]:STORe:CONFiguratio

Description

This command saves a configuration file.

For this command to be accepted, the OSA

module must be in Ready state with no

acquisition in progress.

This command is an event and has no associated

*RST condition or query form.

Syntax

:MMEMory[1..n]:STORe:CONFiguratio

ilename>

Parameter(s)

Filename:
The program data syntax for is

defined as a

element.

Saves a configuration file name.

The parameter can either be only

the filename or the name and its path.

If no path is specified, the default path is used.

The default path name depends on location of

driver libraries.

Example(s)

MMEM:STOR:CONF "config001.cfg"

See Also

MMEMory[1..n]:LOAD:CONFiguration