Managing Results

Optical Spectrum Analyzer

337

Managing Drift Test Results

2. Select a channel from the Channel History tab to view the channel

results for the selected channel.

³

Channel name: indicates the name of the channel as defined in the
channel list.

³

Center wavelength/frequency: indicates the spectral center-of-mass
for the peak in that channel.

³

Signal power: indicates the signal power for the selected channel
(excludes noise).

³

OSNR: is the Optical Signal to Noise Ratio, given by Signal power
(according to the current calculation method, in dBm) minus Noise
(according to the current calculation method, in dBm).

³

Noise: indicates the level of the noise for the selected channel.

³

Bandwidth 3.00 dB: indicates bandwidth measured by taking the
width of a signal at 50 % linear power of the peak or -3 dB from the
peak.