beautypg.com

Calculate[1..n]:channel:add – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual

Page 457

background image

SCPI Command Reference

Optical Spectrum Analyzer

449

Product-Specific Commands—Description

Central wavelenght of the channel being added.

Range is dependent on OSA module limits.

³

Width:

The program data syntax for is defined

as a

element.

Width of the channel being added.

Range: [0.008 … 400.000] nm or [1.0 … 58150.0]

GHz

³

WidthUnit:

The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are: M|HZ.

Width unit of the channel being added.

³

SignalPowerCalculation:

The program data syntax for the second

parameter is defined as a

PROGRAM DATA> element. The allowed

elements for
this parameter are:

INTEGRATEDSIGNAL|PEAKSIGNAL|TOTALCHA

NNEL.
Selects the Signal Power Calculation method to

apply for signal power value.

INTEGRATEDSIGNAL, selects Integrated signal

power method to calculate signal power value.
PEAKSIGNAL, selects Peak signal power method

to calculate signal power value.

TOTALCHANNEL, selectsTotal channel power

method to calculate signal power value.

:CALCulate[1..n]:CHANnel:ADD