beautypg.com

EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual

Page 374

background image

Managing Results

366

FTB-5240S/5240BP

Managing Spectral Transmittance Test Results

³

Channel range (GHz or nm): indicates the limit inside which the
power values will be considered in the channel. Integrated power
is calculated on channel width.

³

Bandwidth 1 at (dB): indicates bandwidth measured by taking the
width of a signal at x dB from the peak.

³

Bandwidth 2 at (dB): indicates bandwidth measured by taking the
width of a signal at y dB from the peak.

³

Input wavelength offset (nm): indicates the offset value applied
on the input wavelength. You can use an offset to adjust your unit.
This does not replace a calibration performed at EXFO, but it can
help you achieve the specifications if you have determined that, for
example, your modules are used beyond the normal allowed.
Entering a value in THz is not possible. The wavelength offset
appears below the graph, on the right, next to the

arrow.

³

Input power offset (dB): indicates the offset value applied on the
input power. You can use an offset to adjust your unit. This does
not replace a calibration performed at EXFO, but it can help you
achieve the specifications if you have determined that, for
example, your modules are used beyond the normal allowed. The
power offset appears in the graph, on the left, above the

arrow.

To edit the power offset value in percentage, press Edit % button.

The percentage value entered in Edit percentage will be
converted to a corresponding equivalent value in dB.