Optical Spectrum Analyzer

499

Product-Specific Commands—Description

:SENSe[1..n]:AVERage:INBand:COUNt?

Description

This query returns number of scans Inband to

produce final trace.

At *RST, this value is 64.

Syntax

:SENSe[1..n]:AVERage:INBand:COUNt?

Parameter(s)

None

Response Syntax

Response(s)

InbandCount:
The response data syntax for is

defined as a

element.

Number of In-Band scans used to average in final

scan.

Example(s)

SENS:AVER:INBand:COUN?

See Also

SENSe[1..n]:AVERage:INBand:COUNt