beautypg.com

EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual

Page 320

background image

Managing Results

312

FTB-5240S/5240BP

Managing WDM Test Results

³

Signal power calculation: indicates which calculation method to
apply for signal power value.

Integrated signal power: The integrated signal power represents
the sum of the power values included between the channel limits
of this channel, minus the estimated noise contribution between
the same boundaries. In some cases, for instance CATV signals,
signals with high-frequency modulation, or signals with an
inherent line width similar or larger than the OSA's resolution
bandwidth, this calculation becomes a better estimation of the
true signal power.

Peak signal power: The peak signal power represents the
maximum power value inside the channel, but it differs a little
from the peak measurement on the spectrum due to the fact that
the estimated noise is subtracted to get the peak signal power.

Total channel power: The total channel power represents the sum
of the power values included inside the channel width (including
noise).

³

Noise for OSNR: indicates which calculation method to use for
OSNR value.

Fixed range IEC based (IEC): The IEC method uses the
interpolation of noise measured on both sides of the signal to
estimate the noise level. The position at which the noise is
estimated from the center wavelength is given by the OSNR
distance.

In-Band (InB): The In-Band method uses a series of scans having
different polarization states to calculate the noise level under the
peak (In-Band).

In-Band narrow filter (InB nf): The In-Band narrow filter method
uses additional processing to provide an accurate OSNR value for
the narrow carved noise. This is because with narrow filters, the
noise level under the peak is not uniform and the OSNR value
depends on the processing width selected.