Managing Results

304

FTB-5240S/5240BP

Managing WDM Test Results

Global Results Tab

The application allows you to view the global results of the current
measurement. You can also edit global parameters to reanalyze the active
trace. The pass/fail verdict for thresholds are displayed in the Global
Results
tab. If the verdict is fail for any parameter, its value will appear in
red. If the verdict is pass, its value will appear in green.

To view global results:
From the main window, select the Global Results tab.

Results for the following parameters for all the channels will be displayed:

³

Average signal power: indicates the sum of the signal powers of
all the peaks detected in the current acquisition, divided by the
total number of peaks.

³

Signal power flatness: indicates the difference between the
maximum and minimum signal power values of the detected
peaks, in dB.

³

Average OSNR: indicates the sum of the entire OSNR of the peaks
detected in the current acquisition, divided by the total number of
peaks.

³

OSNR flatness: indicates the difference between the maximum
and minimum OSNR values of the detected peaks, in dB.