Preparing Your Optical Spectrum Analyzer for a Test
24
FTB-5240S/5240BP
Performing User Calibration
A high level sequence of steps to perform user calibration are listed below,
with the detailed instructions in the following procedures.
³
Proper warm-up is required before performing user calibration.
³
If user correction factors are present in the module (indicated by a
message in calibration page),
³
save a backup copy of the correction factors (optional but strongly
recommended).
³
clear user correction factors.
³
Make measurements.
³
Write measurement to text file according to following format.
³
First column: reference wavelength in nm
³
Second column: wavelength (in nm) read using the OSA
application and module
³
Third column: reference power in dBm
³
Fourth column: power (in dBm) read using the OSA application
and module
100 calibration points limit in the application. Each value in the file
must be separated using semi-colon (;) character.
³
Write new correction factors inside the module
³
Load the text file built at preceding step
³
Press Write to Module.
The application shows a warning message indicating to clear
correction factors before making new measurements
³
Verify that calibration points are properly applied.