Setting Up the Instrument in Drift Mode

Optical Spectrum Analyzer

163

Setting Up Acquisition Parameters

3. Select a sampling unit and enter a sampling count to configure a time

that should be elapsed between the start of each acquisition during a
drift measurement.

4. Select a duration unit and enter a duration count to configure the total

duration of a drift measurement.

5. Enter a drift file name that should be used to save a drift file.

6. Select a location where the drift file should be saved.