SECTION 15 9600 SPECIFICATIONS
15.1 SENSOR SPECIFICATIONS
USING 2 MM PATHLENGTH DETECTOR CELL
Range: 0 to 10 mg/L Nitrate-N
Accuracy: 0-6 mg/L -- +/- 5 % of reading or 0.2 mg/L, whichever is greater.
Accuracy: 6-10 mg/L -- +/- 10 % of reading using a 6 mg/L Nitrate-N standard. Accuracy in the
0-6 mg/L range may be slightly compromised by the use of the 6 mg/L standard.
Detection Limit: 0.025 mg/L Nitrate-N
USING 10 MM PATHLENGTH DETECTOR CELL
Range: 0 to 2 mg/L Nitrate-N
Accuracy:
+/- 5 % of reading or 0.02 mg/L, whichever is greater
Detection
Limit:
0.005 mg/L
15.2 SYSTEM SPECIFICATIONS
Operating Temperature Range: 1 to 45 C
Operating Depth : < 200 feet (61 m) of water
Reagent Life: > 30 days at a 1 hour sample interval
Standard Battery Configuration: YSI 9611 Battery Pack with 14 Alkaline D Cells.
Battery Life: Approx. 70 days at a 1 hour sample interval at 20 C; approximately 30 days at a 1 hour
sample interval at 5 C.
Dimensions: Cylindrical with a diameter of 13 inches and a height of 28 inches.
Weight With All Reagent Bags Full and Alkaline Battery Pack Installed: Approximately 40 lbs.
(18.2 kg).
Communication Protocol:
RS-232 for communication to PC
SDI-12 for communication to Data Collection Platforms
YSI Environmental
86
Model 9600 Nitrate Monitor