Packing instructions – YSI 9600 User Manual
Page 84
• Disconnect the tubing from the cadmium column assembly and plug the ends with the two black
plugs that were set aside during unpacking. Connect the two nuts that were removed from the
cadmium column with the translucent coupling nut that was set aside during unpacking. Do
NOT return the cadmium column assembly with the analyzer.
• Fill the second standard reagent bag with purified water and attach to the sample inlet fitting via
the tubing assembly.
• Locate the six reagent lines (previously attached to reagent bags) and place them into a vessel
that contains approximately 500 mL of purified water.
• Connect to the 9600 via a computer running NUview software and run the predeployment test to
assure that the system is flushed thoroughly.
• Remove the second standard bag from the inlet port and replace with the inlet filter and its
guard. Remove the reagent lines from the water vessel and allow to hang down into the lower
compartment of the carousel.
• Close the pinch clamps on the waste bags and remove them from the carousel. Do NOT return
the waste bags with the instrument.
• Reassemble the 9600 by placing the carousel into the outer shell and attaching the four assembly
screws.
PACKING INSTRUCTIONS
) Clean and decontaminate items to insure the safety of the handler.
) Complete and include the Cleaning Certificate.
) If possible, place the product in a plastic bag to keep out dirt and packing material.
) Use a large carton, preferably the original, and surround the product completely with packing material.
) Insure for the replacement value of the product.
Cleaning Certificate
Organization
Department
Address
City
State Zip
Country
Phone
Model No. of Device
Lot Number
Contaminant (if known)
Cleaning Agent(s) used
Radioactive Decontamination Certified?
(Answer only if there has been radioactive exposure) Yes
No
Cleaning Certified By
Name
Date
YSI Environmental
84
Model 9600 Nitrate Monitor