beautypg.com

Off prs, Yes prs sens, 00 ref na 0 – TeeJet 854 Sprayer Control User Manual

Page 21: 854 sprayer control

background image

www.teejet.com

16

854 Sprayer Control

Figure 4-7: Pressure Hold

Off

PrS

Pressure.Sensor.Installed

NOTE: If “No” is selected in this step, the next

two screens/steps in this User Guide will

not be displayed on the console during

programming.

The Pressure Sensor Installed step indicates to the

console whether or not a pressure sensor has been

installed on the sprayer.

• Use the PLUS or MINUS keys to select

“Yes” or “No”.

• Select “Yes” if a pressure transducer is

installed.

• Select “No” if a pressure transducer is not being

used.

• Press the PROGRAM key to accept the

value and advance to the next step.

Figure 4-8: Pressure Sensor Installed

YES

PrS

SEnS

Pressure.Transducer.Low.

Pressure.Calibration.(P.Ref)

This function is used to calibrate the Zero Pressure

setting of the pressure transducer installed on the

system. The pressure transducer used with the 854

is a current type transducer and uses a 4-20 mA

reading. “4.0 mA” represents “Zero” pressure.
Figure 4-9: Pressure Transducer

4.00

rEF

nA

0

PSI

Automatic Calibration

• Make sure that the sprayer pump is turned off

and that there is absolutely no pressure in the

system (release pressure held by boom control

valves and nozzle body check valves).

• In some cases it may be best to remove the

sensor from the plumbing system to complete

the calibration.

• Press and hold the PLUS and MINUS

keys simultaneously for 3 seconds to activate

the automatic calibration feature.

• The lower left portion of the display will count

from 1 - 10 during the calibration.

• Once the display finishes counting, it should

display a number close to 4.0 (+/- 0.2).

• The low pressure value of the transducer has

been calibrated.

• Press the PROGRAM key to advance to the

next step.