beautypg.com

Dangers, Warning – Reznor PDH Operation Manual PreevA User Manual

Page 21

background image

Form O-PreevA, P/N 234661R8, Page 21

3) Remove

Cover Plate

2) Remove

grommet.

1) Disconnect

the drain.

4) Slide out

the drain pan.

3.7.3 Optional Ultra-

Violet Light

FIGURE 8 -

Slide Out

Drain Pan

NOTE: Indoor

unit is illustrated.

Outdoor unit is

the same except

drain is in the

bottom panel

not the base.

3.7.4 Optional Reheat Module

Reheat Module Coils

- Follow the instructions in Paragraph 3.7.1 to clean the reheat module coils.

Reheat Module Compressor Maintenance and Replacement

Compressor Handling

Do not lift compressor by copper tubing. To prevent internal damage, compressors

must ALWAYS be held upright.

DANGERS

The reheat circuit is a high pressure system. Hazards exist that could result in

personal injury or death. It is therefore required that the removal and installation of

this scroll compressor be performed by a technician qualified in R410A refrigerant.

See Hazard Levels, page 2.

Never use oxygen to pressurize a refrigeration system. Oxygen can explode on

contact with oil and could result in personal injury or death. When using high

pressure gas such as nitrogen for this purpose, ALWAYS USE A PRESSURE

REGULATOR that can control the pressure down to 1 or 2 psig. Failure to use

a regulator will result in extremely high pressure which could exceed the burst

pressure of the compressor or other system components and result in personal

injury or death. See Hazard Levels, page 2.

WARNING

For your safety, wear eye protection, gloves, and protective clothing when handling

refrigerant and oil and when brazing. Have a fire extinguisher nearby. See Hazard

Levels, page 2.

The ultra-violet light is located in the blower compartment. It has its own power supply.

Disconnect all power before performing maintenance or service.
Replace the bulb(s) periodically to maintain design specifications Replace the bulb(s)

after 6000 hours of operation or one year, whichever comes first. For best perfor-

mance, replace the bulb(s) at the beginning of each cooling season.
The bulb(s) should be light blue when lit. If the bulb color changes to red or flickers,

replace the bulb. Output can be checked with a UVC radiometer. Replace bulb when

output falls to 70% of initial reading.
Use factory-authorized replacement bulb only. Other bulbs may plug into the fixture

but could cause overheating and should not be used.

algae formation in the drain

pan and traps. The drain

pan and traps must be

cleaned regularly to avoid

blockage that can slow or

stop water flow, resulting in

backup into the cabinet.

WARNING

DO NOT use ultra-

violet light for

service lighting.

Never expose eyes

or skin to ultra-

violet light.