Caution: Welch Allyn recommends the use of the Single Use Insufflation Bulb

Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate

the need to properly process the insufflation bulb with each use, but restricts internal

contamination and aspiration of fluids.
Caution: Do not disassemble the insufflation bulb.
Caution: The insufflation bulb may discolor after disinfection or sterilization

without effecting performance.
Caution: Do not autoclave insufflation system. Damage will occur.

1. Prepare an enzymatic detergent formulated for endoscopic instruments

according to the manufacturer’s instructions.These instructions were

validated using ENZOL/CIDEZYME.

2. Submerge the item in the cleaning solution

3. Using a soft bristle brush, scrub the item submerged in the cleaning

solution for a minimum of 5 minutes then rinse with sterile water for

5 minutes.

4. Using a 60 cc syringe, deliver 60 cc of the cleaning solution to the inside

of the bulb through the hose-end.

5. Lightly agitate the solution inside of the bulb for a minimum of 5 minutes.

6. Empty the cleaning solution from the bulb.

7. Using a 60 cc syringe, deliver 60 cc of sterile water into the bulb through

the hose-end.

8. Lightly agitate the water inside of the bulb for a minimum of 5 minutes.

9. Empty the rinse water from the bulb, and repeat rinse with sterile water

5 more times.

10. The bulb must be dried sufficiently. Any moisture inside the device could

occlude the insufflation bulb filter (if used) or dilute disinfection or

sterilization agents. One way to achieve this is to hang the bulb with the

tube down for a minimum of 24 hours.

Insufflation System

Do not disassemble

this connection

Part Number: 30200

Cleaning

5