Insufflation system, Cleaning – Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes, Anoscopes and Accessories - User Manual User Manual
Page 7

Caution: Welch Allyn recommends the use of the Single Use Insufflation Bulb
Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate
the need to properly process the insufflation bulb with each use, but restricts internal
contamination and aspiration of fluids.
Caution: Do not disassemble the insufflation bulb.
Caution: The insufflation bulb may discolor after disinfection or sterilization
without effecting performance.
Caution: Do not autoclave insufflation system. Damage will occur.
1. Prepare an enzymatic detergent formulated for endoscopic instruments
according to the manufacturer’s instructions.These instructions were
validated using ENZOL/CIDEZYME.
2. Submerge the item in the cleaning solution
3. Using a soft bristle brush, scrub the item submerged in the cleaning
solution for a minimum of 5 minutes then rinse with sterile water for
5 minutes.
4. Using a 60 cc syringe, deliver 60 cc of the cleaning solution to the inside
of the bulb through the hose-end.
5. Lightly agitate the solution inside of the bulb for a minimum of 5 minutes.
6. Empty the cleaning solution from the bulb.
7. Using a 60 cc syringe, deliver 60 cc of sterile water into the bulb through
the hose-end.
8. Lightly agitate the water inside of the bulb for a minimum of 5 minutes.
9. Empty the rinse water from the bulb, and repeat rinse with sterile water
5 more times.
10. The bulb must be dried sufficiently. Any moisture inside the device could
occlude the insufflation bulb filter (if used) or dilute disinfection or
sterilization agents. One way to achieve this is to hang the bulb with the
tube down for a minimum of 24 hours.
Insufflation System
Do not disassemble
this connection
Part Number: 30200
Cleaning
5