Rigid sigmoidoscopes, anoscopes and light carrier, Cleaning – Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes, Anoscopes and Accessories - User Manual User Manual

Page 4

background image

Caution: Clean and disinfect all devices or accessories per these instructions prior

to returning to Welch Allyn (e.g. for repair and service) .

Caution: Users are responsible to qualify any deviations from these procedures.

Rigid Sigmoidoscopes, Anoscopes and Light Carrier

Anoscope

Part Numbers:

37019 38614

37023 38619

37027 38622

38108 38900

38114 39614

38119 39619

38122 39622

Sigmoidoscope

Part Numbers:

32010 32820

32020 32830

32410 33220

32420 33620

32810 33830

Light Carrier

Part Numbers:

36019, 38700
Suction Tube*

Part Numbers:

30130, 30140

* Suction Tubes not illustrated

NOTE For all steps, remove obturator for separate cleaning and open

viewing window.

Caution: Welch Allyn recommends the use of the Single Use Insufflation Bulb

Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate

the need to properly process the insufflation bulb with each use, but restricts internal

contamination and aspiration of fluids.

Cleaning

1. Prepare an enzymatic detergent formulated for endoscopic instruments

according to the manufacturer’s instructions. These instructions were

validated using ENZOL®/CIDEZYME®.

2. Submerge the item in the cleaning solution.

3. Using a soft bristle brush, scrub the item submerged in the cleaning solution

for a minimum of 5 minutes then rinse with sterile water for 5 minutes.

37023

32820

36019