Caution: Clean and disinfect all devices or accessories per these instructions prior
to returning to Welch Allyn (e.g. for repair and service) .
Caution: Users are responsible to qualify any deviations from these procedures.
Rigid Sigmoidoscopes, Anoscopes and Light Carrier
Anoscope
Part Numbers:
37019 38614
37023 38619
37027 38622
38108 38900
38114 39614
38119 39619
38122 39622
Sigmoidoscope
Part Numbers:
32010 32820
32020 32830
32410 33220
32420 33620
32810 33830
Light Carrier
Part Numbers:
36019, 38700
Suction Tube*
Part Numbers:
30130, 30140
* Suction Tubes not illustrated
NOTE For all steps, remove obturator for separate cleaning and open
viewing window.
Caution: Welch Allyn recommends the use of the Single Use Insufflation Bulb
Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate
the need to properly process the insufflation bulb with each use, but restricts internal
contamination and aspiration of fluids.
Cleaning
1. Prepare an enzymatic detergent formulated for endoscopic instruments
according to the manufacturer’s instructions. These instructions were
validated using ENZOL®/CIDEZYME®.
2. Submerge the item in the cleaning solution.
3. Using a soft bristle brush, scrub the item submerged in the cleaning solution
for a minimum of 5 minutes then rinse with sterile water for 5 minutes.
37023
32820
36019