beautypg.com

Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes, Anoscopes and Accessories - User Manual User Manual

Page 5

background image

1. Follow the cleaning instructions above.

2. Using CIDEX OPA (0.55% OPA, 14-day maximum reuse), soak the item for

a minimum of 12 minutes.

3. Perform 3 sequential rinses of the item for a minimum of 5 minutes each

with sterile water.

4. Allow the item to dry for 24 hours or until no moisture remains in the device.

Any moisture in the device could occlude the insufflation bulb filter.

High Level Disinfection

Sterilization

Light Handle Assembly and Handle Adapter

Assembly and Disassembly Instructions for 73210

and 73211

1. Follow the Cleaning instructions above.

2. Perform one autoclave cycle that includes 6 minutes at 270° F (132° C)

and a cool down period of at least 15 minutes. These instructions were

validated using a gravity autoclave. Validate any other sterilization

equipment and load configurations.

Part Numbers 73210 and 73211

(including 07800-U lamp)

Part Number 73500

(including 03100-U lamp)

Warning: Allow the lamp to cool for at least 5 minutes before handling

Note: Leave the lamp installed while reprocessing 73210, 73211 and 73500.

1. Disconnect the light handle from the power supply cord and the device.

2. Unscrew the outer sleeve of the handle by turning it counterclockwise.

Rectal Light Handle Body and

Cord

Rectal Light Handle

Outer Sleeve

CCW

CW

3