Caution: Do not saturate the rheostat area. Excess solution entering the assembly

may damage internal components in the handle.

Rechargeable Handle

Part Number: 71000-A

Rheostat

(Brightness Control)

1. Using CaviWipes, wipe the surface to remove all gross debris.

2. Using a soft bristle brush, scrub the handle for a minimum of 5 minutes then

repeat wipe with a second CaviWipe.

3. Allow the handle surface to dry for a minimum of 10 minutes.

1. Follow the Cleaning instructions above.

2. Using CaviWipes, wipe the handle and keep wet with CaviWipe solution for

a minimum of 5 minutes.

3. Allow the handle surface to dry for a minimum of 10 minutes.

Cleaning

Intermediate Level Disinfection

1. Prepare an enzymatic detergent formulated for endoscopic instruments

according to the manufacturer’s instructions. These instructions were

validated using ENZOL/CIDEZYME.

2. Wet the cord with the cleaning solution.

3. Using a soft bristle brush, scrub the cord with the cleaning solution for a

minimum of 5 minutes then rinse the cord with sterile water for 5 minutes.

1. Follow the Cleaning instructions above.

2. Using CaviWipes, wipe the cord and keep wet with CaviWipe solution for

a minimum of 5 minutes.

3. Allow the cord surface to dry for a minimum of 10 minutes.

The following cleaning instructions apply to the Power Cord ONLY:

1. Disconnect the power supply cord from the light handle.

2. Disconnect the power supply from its wall receptacle.

Cleaning the Power Cord

Intermediate Level Disinfection of the Power Cord

8