Localize Station settings

Welch Allyn

Auto discharge settings for confirmed patients

1.

Settings starting with “AUTODISCHARGE” control automatic discharge behavior for
confirmed patients.

Figure 12-14: Confirmed patients auto discharge related settings

2.

The following table describes the function of each setting.

Table 12-2: AUTODISCHARGE setting functions and default values

Setting Label

Function

Default value

ENABLED

Turns Auto Discharge functions on/off for confirmed
patients

TRUE (on)

EXECUTIONFREQUENCY.MINUTES

Controls how often the process runs, in MINUTES

60

EXECUTIONSEQUENCE

Controls the run order between Late ADT Matching,
Confirmed Auto Discharger, and ADT Auto Discharger
processes.

2 (second)

UNLINKED.STALEPERIOD.MINUTES

Determines a period of time in MINUTES until
automatic discharge of a patient for which no activity
has taken place for the patient including ADT
messages, Continuous Monitoring, Spot Vitals taken,
edits to demographics or visit data.

1440 (24 hours)

ADTLINKED.STALEPERIOD.DISCHARGE.

MINUTES

Determines a period of time in MINUTES until
automatic discharge of a patient after:

receipt of an ADT discharge message, and

time of the last Spot Vitals taken or Continuous
Monitoring sessions ends.

240 (4 hours)

ADTLINKED.STALEPERIOD.TRANSFER.

MINUTES

Determines a period of time in MINUTES until
automatic discharge of a patient after:

receipt of an ADT transfer message, and

time of the last Spot Vitals taken or Continuous
Monitoring sessions ends.

240 (4 hours)

Caution Do not change settings for EXECUTION SEQUENCE unless directed
by Welch Allyn Engineering.