Specifications
Welch Allyn Connectivity Server
Realtime page
Details page
Table 15. WACS AcuityLink option: Clinician Notifier Realtime page
Feature
Specification
Realtime waveforms available for display
ECG, SpO
2
, respiration, CO
2
, invasive pressure
Time before a realtime waveform appears in Clinician
Notifier
2-3s
Amount of waveform data displayed
Up to 9s, depending upon mm/sec setting
Source lead shown upon opening the page
II
Default amplitude
1X
Default sweep speed
Respiration and CO
2
: 6.25 mm/s
ECG, invasive pressure (P1, P2) and SpO
2
: 25 mm/s
Waveform shown upon opening the page
If showing a moment in time when the patient is (or was)
alarming, the waveform is the waveform that caused the
alarm.
Otherwise, the waveform is ECG.
Table 16. WACS AcuityLink option: Clinician Notifier Details page
Feature
Specification
Amount of waveform data displayed
Up to 9s, depending upon mm/sec setting
Source lead shown upon opening the page
II
Default amplitude
1X
Waveform shown upon opening the page
If showing a moment in time when the patient is (or was)
alarming, the waveform is the waveform that caused the
alarm.
Otherwise, the waveform is ECG.
Length of time alarm histories are shown on mobile
devices after patient monitor is disconnected from
Acuity System
Default 96 hours
Configurable: 1-96 hours, but no longer than the Acuity
System Full Disclosure license period