39
Multiple patient observations setting on
If this setting is checked, WACS sends all vital signs from all patients in a single HL7
message.
In this example, one message includes these components:
•
Data for patient Van Goe and patient Bizmark (in bold)
•
One acknowledgement
Message resulting from multiple patient observations setting on
HL7# HL7Log Initiator sending message:
MSH|^~\&|WAP^WAP||||20090129091001.081-0800||ORU^R01|200901290910010813f8|P|2.4
PID|1|9582173|9582173||Van Goe^Edgar^A|||||||||||||9582173
PV1|1||TEST^102^GA001150^^^^^^PropaqCS 246
OBR|1|||VITALS^Vital Signs^WAP|||20090129090800.000-0800
OBX|1|ST|Heart Rate^Heart Rate^WAP||80|^BPM||||||||20090129090800.000-0800
OBX|2|ST|Temperature^Temperature^WAP|1|98.6|^°F||||||||20090129090800.000-0800
OBX|3|ST|Temperature^Temperature^WAP|2|97.5|^°F||||||||20090129090800.000-0800
OBX|4|ST|SPO2^SPO2^WAP||97|^%||||||||20090129090800.000-0800
OBX|5|ST|CO2 (In)^CO2 (In)^WAP||0.0|^%||||||||20090129090800.000-0800
OBX|6|ST|CO2 (Ex)^CO2 (Ex)^WAP||4.9|^%||||||||20090129090800.000-0800
OBX|7|ST|RR/BR^RR/BR^WAP||12|^Br/M||||||||20090129090800.000-0800
OBX|8|ST|PVC^PVC^WAP||0|^PVC/Min||||||||20090129090800.000-0800
PID|2|1567252|1567252||Bizmark^Michelle^D|||||||||||||1567252
PV1|2||TEST^103^GA005267^^^^^^PropaqCS 242
OBR|1|||VITALS^Vital Signs^WAP|||20090129090800.000-0800
OBX|1|ST|Heart Rate^Heart Rate^WAP||80|^BPM||||||||20090129090800.000-0800
OBX|2|ST|Temperature^Temperature^WAP|1|98.6|^°F||||||||20090129090800.000-0800
OBX|3|ST|Temperature^Temperature^WAP|2|97.5|^°F||||||||20090129090800.000-0800
OBX|4|ST|SPO2^SPO2^WAP||97|^%||||||||20090129090800.000-0800
OBX|5|ST|RR/BR^RR/BR^WAP||12|^Br/M||||||||20090129090800.000-0800
OBX|6|ST|PVC^PVC^WAP||0|^PVC/Min||||||||20090129090809.000-0800
HL7# HL7Log Initiator received message:
MSH|^~\&|||WAP^WAP||20090129091001.081-0800||ACK^R01|200901290910010813f8|P|2.4
MSA|AA|200901290910010813f8