AcuityLink option

Compatibility

General capabilities and settings

Alarm and alert indications

Table 7. WACS AcuityLink option: compatibility

AcuityLink Clinician Notifier software option
to WACS

Compatibility

Clinician Notifier software version 1.4x

Welch Allyn Connectivity Server software version 2.6x

Acuity Central Monitoring System software version 8.2x

Table 8. WACS AcuityLink option: general capabilities and settings

Feature

Specification

Number of minutes after dropout before automatic
logout

Default: 3 min
Configurable: 3-30 min

Password required to enter Clinician Notifier
program on mobile device

Default: required
Configurable: required or not required

Clinician Notifier software time synchronization

Synchronized to the WACS server via Network Time Protocol (NTP).

Synchronized to Acuity System and the WACS server, within 2s

Table 9. WACS AcuityLink option: Clinician Notifier alarm and alert indications

Feature

Specification

Visual

Alarm popup window
Single alarm popup: a patient’s highest priority alarm only
Multiple alarm popup: multiple patients in alarm

Alarm status bar message
Number of alarms (only one per patient is counted)
Number of suspended alarms

Red LED light, if available on mobile device

Time format

24-hour

Auditory/vibratory notification

Default: audio and vibration

Configurable at WACS: audio, vibration or both
User cannot use mobile device OS audio control icon to
override WACS setting to turn on or off vibration or audio for
Clinician Notifier alarm delivery.

If WACS specifies vibration only and devices do not have
vibratory capability, device defaults to audio with 20%
volume level.

If WACS specifies audio, mobile devices always deliver at
least 20% volume level, regardless of OS volume setting on
device.