Acuitylink option – Welch Allyn Connectivity Server - User Manual User Manual
Page 87

Directions for Use
Specifications
83
AcuityLink option
Compatibility
General capabilities and settings
Alarm and alert indications
Table 7. WACS AcuityLink option: compatibility
AcuityLink Clinician Notifier software option
to WACS
Compatibility
Clinician Notifier software version 1.4x
Welch Allyn Connectivity Server software version 2.6x
Acuity Central Monitoring System software version 8.2x
Table 8. WACS AcuityLink option: general capabilities and settings
Feature
Specification
Number of minutes after dropout before automatic
logout
Default: 3 min
Configurable: 3-30 min
Password required to enter Clinician Notifier
program on mobile device
Default: required
Configurable: required or not required
Clinician Notifier software time synchronization
Synchronized to the WACS server via Network Time Protocol (NTP).
Synchronized to Acuity System and the WACS server, within 2s
Table 9. WACS AcuityLink option: Clinician Notifier alarm and alert indications
Feature
Specification
Visual
•
Alarm popup window
Single alarm popup: a patient’s highest priority alarm only
Multiple alarm popup: multiple patients in alarm
•
Alarm status bar message
Number of alarms (only one per patient is counted)
Number of suspended alarms
•
Red LED light, if available on mobile device
Time format
24-hour
Auditory/vibratory notification
•
Default: audio and vibration
•
Configurable at WACS: audio, vibration or both
User cannot use mobile device OS audio control icon to
override WACS setting to turn on or off vibration or audio for
Clinician Notifier alarm delivery.
If WACS specifies vibration only and devices do not have
vibratory capability, device defaults to audio with 20%
volume level.
If WACS specifies audio, mobile devices always deliver at
least 20% volume level, regardless of OS volume setting on
device.